Precision Medical’s low-flow and high-flow air-oxygen blenders offer unparalleled flexibility, safety, and value. Our innovative, modular systems can be configured to suit various clinical needs within minutes.
The PM5300 series of low-flow air-oxygen blenders is ideal for NICU and PICU settings. This series delivers FiO2 mixtures from two outlet ports and has an overall flow range of 0-30 l/min. The PM5300 series is a highly accurate unit that maintains FiO2 even at low flows.
The PM5200 series of high-flow air-oxygen blenders is intended for use in ER departments, adult ICUs, ORs, and labor and delivery settings. It delivers FiO2 mixtures from two outlet ports and has an overall flow range of 2-120 l/min.
Clinicians have the option to turn the bleed control on or off which provides quieter operation. A large selection control knob is recessed to prevent accidental damage and makes oxygen percentage adjustments straightforward.
The flexible design and available accessories allow for customization. Manage your capabilities as demand and budget warrant. Extend the base blender setup by adding the optional Oxygen Monitor or AccuO2 Oxygen Analyzer to create a complete mixing package.
The Air-Oxygen Blender is easily installed with a universal mounting bracket and can be quickly detached without removing hoses. A unique, interchangeable design reduces routine maintenance time and costs. The infection-control friendly, ABS housing makes the blender durable and effortless to clean. The unit has 8 simple-to-replace components and is easy to service by biomed departments.
For custom outlet configurations and assembly sizes, please contact customer service by using this form or by calling 1-800-272-7285.
Caution! U.S. Federal law restricts this device to sale by or on the order of a physician.
PM5200 High Flow
15 – 120 l/min
supply pressures at 50 psi (3.4 bar) with BLEED closedPM5300 Low Flow
3 – 30 l/min
supply pressures at 50 psi (3.4 bar) with BLEED closed
PM5200 High Flow
20 – 100 l/min
supply pressures at 50 psi (3.4 bar) with BLEED closedPM5300 Low Flow
0 – 30 l/min
supply pressures at 50 psi (3.4 bar) with BLEED closed
PM5200 High Flow
13 l/min or less at 50 psi (3.4 bar)PM5300 Low Flow
3 l/min or less at 50 psi (3.4)
PM5200 High Flow
≥ 120 l/minPM5300 Low Flow
≥ 30 l/min
(Loss of Air or Oxygen Supply)
PM5200 High Flow
≥ 85 l/minPM5300 Low Flow
≥ 45 l/min
PM5200 High Flow
50 psi(3.45 bar)
13 – 25 psi (0.9 – 1.7 bar)PM5200 High Flow
60 psi(4.14 bar)
16 – 24 psi (1.1 – 1.65 bar)
PM5300 Low Flow
50 psi(3.45 bar)
18 – 22 psi (1.2 – 1.5 bar)
PM5300 Low Flow
60 psi(4.14 bar)
16 -24 psi (1.1 – 1.65 bar)
When pressure differential is 6 psi (0.4 bar) or less
≥ to 80 db at 1 ft (0.3 m)
21 – 100%
30 – 75 psi (2.1 – 5.2 bar) Air and Oxygen
within 10 psi (0.69 bar) of each other
± 1% Oxygen
DISS Type – Air & Oxygen Inlets & Outlets and/or NIST Type – Air & Oxygen Inlets
Depth: 4.9 in (12.5 cm)
Width: 2.3 in (5.7 cm)
Height: 4.1 in (10.4 cm)
Unit Weight: 2.29 lbs (1.04 kg)
Shipping Weight: 2.95 lbs (1.34 kg)
59°F to 104°F (15°C to 40°C)
Temperature Range: -10°F to 140°F (-23°C to 60°C)Humidity: Max 95% Non-condensing
± 3% of full scale
Low Flow: ≤ 2 psi (0.14 bar) at inlet pressure from 30 – 90 psi (2.1 – 6.2 bar)
and at 10 l/min flow rate at 60% FI02High Flow: ≤ 3 psi (0.21 bar) at inlet pressure from 30 – 90 psi (2.1 – 6.2 bar)
and at 30 l/min flow rate at 60% FI02
The Air-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 21647.
Air: Medical Air supply should meet the requirements of ANSI Z86.1 – 1973 commodity specification for Air, type 1 grade D or better.Oxygen: Oxygen supply must meet all requirements of USP Medical Grade Oxygen.Dew Point(ONLY for CE requirements): Both inlets should remain 10°F (5.55°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.
Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 l/min for the high flow Blender, and 3 l/min for the low flow Blender.
Manuals
- PM52/5300 Air-Oxygen Blender User Manual – English
- PM52/5300 Air-Oxygen Blender User Manual – Danish
- PM52/5300 Air-Oxygen Blender User Manual – Dutch
- PM52/5300 Air-Oxygen Blender User Manual – Finnish
- PM52/5300 Air-Oxygen Blender User Manual – French
- PM52/5300 Air-Oxygen Blender User Manual – German
- PM52/5300 Air-Oxygen Blender User Manual – Greek
- PM52/5300 Air-Oxygen Blender User Manual – Irish
- PM52/5300 Air-Oxygen Blender User Manual – Italian
- PM52/5300 Air-Oxygen Blender User Manual – Spanish
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that the Blender, (the Product), will be free of defects in workmanship and/or material for the following period:
Two (2) years from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.